Fluorescence hydrogen peroxide probe based on titanium dioxide film modified microstructured polymer optical fiber
来源:王丽莉研究员个人网站 发布日期:2010-07-27
作者:Dongdong Li, Lili Wang
关键字:MPOF
论文来源:期刊
发表时间:2010年

A highly sensitive microstructured polymer optical fiber (MPOF) probe for hydrogen peroxide was made by forming a rhodamine 6G-doped titanium dioxide film on the side walls of array holes in an MPOF. It was found that hydrogen peroxide only has a response to the MPOF probe in a certain concentration of potassium iodide in sulfuric acid solution. The calibration graph of fluorescence intensity versus hydrogen peroxide concentration is linear in the range of 1.6 × 10−7 mol/L to 9.6 × 10−5 mol/L. The method, with high sensitivity and a wide linear range, has been applied to the determination of trace amounts of hydrogen peroxide in a few real samples, such as rain water and contact lens disinfectant, with satisfactory results.